Pular para o conteúdo principal

ETHNOS_APP

Início • Busca • Periódicos • Lista 0

Technical report

The effects of humidity on the Abawerk osteometric board

Dados Bibliográficos

ID8318005
AutoresMarie Clabeaux Geise (Buffalo State University, autor correspondente)
Ano1986
Volume71
Fascículo4
Páginas485-486
Data de publicação1986-12-01
Peer ReviewedSim
Open AccessSim
TipoARTICLE
PeriódicoAmerican Journal of Physical Anthropology (JOURNAL)
Identificadores do periódicoISSN: 0002-9483 • E-ISSN: 1096-8644
EditoraWiley (PUBLISHER • GB)
DOI10.1002/ajpa.1330710412
PMID3812663
OpenAlexW2007890900
IdiomaEN
Citações recebidas1
Referências citadas2

Calibration of the Abawerk osteometric board was checked in varying conditions of relative humidity. The paper scale of the board expands in high relative humidity. Measuring errors as great as 3.5 mm may result when using the board in humid environments

Geography · Humidity · Meteorology · Environmental Science · Infrared Thermography in Medicine

  • Investigating the Reliability and Validity of the Portable Osteometric Device

    Open Access•Eric Gary Anderson, Sierra W Mali et al.•Advances in Archaeological Practice•2024

  • Humidity as a source of measurement error in osteometrics

    Open Access•Gene H Albrecht•American Journal of Physical…•1983

  • Multiple observers, humidity, and choice of precision statistics

    Open Access•Charles J Utermohle, Stephen L Zegura et al.•American Journal of Physical…•1983

Obras citantes distintas1
Citações por ano0,5
Intervalo de citações2024 - 2024 (1)
Velocidade de citaçãorecent
Altamente citadoNão
Tipos de citaçãoNeutras: 1
Ethnos_APP • Projeto Open Source • Licença MIT • Frontend v2.0.0 • Privacidade e Cookies • Documentação da API: api.ethnos.app/docs • Código da API: GitHub • DOI: 10.5281/zenodo.17049435 • Código do Frontend: GitHub • DOI: 10.5281/zenodo.17050053 • cruz.rio.br • Expectantes Misericordiae